Testing: I tested this unit using a 115 V (as measured at the time) home power source. When set at 1000V, the unit produced 456 V DC output. Adjusted for the voltage difference, 456 * 250/115 = 991V. Output power goes into standard BNC connector. The unit has a 2.5 amp fuse (you would need to adjust it for a 115V network).
Note: the unit pictured here is the one I already sold. The unit that is being sold here is pretty much identical and was tested with same result.
:SATISFACTION:
Fluke 415B 3.1 kV @ 30mA Power Supply
General Description
The Model 415B High Voltage Power Supply is a stable well regulated source of dc voltage. The Model 415B is capable of providing 0 to 3100 volts dc and 0 to 30 mA current. The output voltage is adjustable in steps of 500, 100, 10 and 1 volt by front panel step switch controls. A vernier control provides a means of fine adjustment within the one volt steps. Output power is available from both a front and rear panel high voltage connector, type UG931/U. Output voltage polarity may be switched by a front panel control to provide either a positive or negative output with respect to chassis ground of the instrument.
The power supply is provided with an overcurrent protection circuit to prevent damage to the instrument in the event of overload or accidental short circuit at the output. Normal operation is quickly restored by a manual reset control following removal of the overload.
The Model 415B combines the best features of electron tubes and solid state components in a hybrid circuit. The improved reliability and temperature stability characteristics of silicon transistor amplifiers and rectifier diodes together with the high voltage handling characteristics of vacuum tubes provide a dependable, low noise level, highly regulated output. Applications of the Model 415B include calibration of voltmeters and powering of photomultiplier tubes, high voltage klystrons or other devices with current requirements up to 30 mA.
ELECTRICAL SPECIFICATIONS
OUTPUT VOLTAGE
Adjustable from 0 to +/-3100 volts dc
OUTPUT CURRENT
0 to 30 mA
OUTPUT POLARITY
Positive or negative with respect to ground (the polarity is controlled by a front panel switch)
LOAD REGULATION
0.0005% or 5 mV (whichever is greater) from a no load to a full load change in output current
LINE REGULATION
0.0005% or 2 mV (whichever is greater) for a 10% line change from nominal
STABILITY
+/-0.0020% per hour; +/-0.01% per day after warmup period and constant line voltage, constant load and constant temperature
RIPPLE
Less than 100 uV RMS; less than 1 mV peak-to-peak
RESOLUTION
5 mV
CALIBRATION ACCURACY
+/-0. 25% or 100 mV (whichever is greater) with. vernier at zero
RESETABILITY
+/-0. 05% or 50 mV
RECOVERY TIME
Within 50 uSec
OVERCURRENT PROTECTION
Adjusted to disconnect output at nominally 30 mA +/- 10% load current. Manual reset
WARMUP TIME
60 minutes to meet specifications.
INPUT POWER
115/230 volts ac +/-100%, 50 to 60 Hz, approximately 300 VA at full load output
MECHANICAL SPECIFICATIONS
OUTPUT CONNECTORS
One UG931/U front and rear (one mating connector supplied)
METER
3100-0-3100 vdc (+/-3%)
WEIGHT
Approximately 30 pounds
ENVIRONMENTAL SPECIFICATIONS
OPERATING TEMPERATURE
0' C to 50' C
STORAGE TEMPERATURE
-20' C to 70' C
HUMIDITY
0 to 80%
ALTITUDE OF OPERATION
Sea level to 10,000 feet
STORAGE OR SHIPPING ALTITUDE
Sea level to 50,000 feet
VIBRATION
Meets MIL-T-945A
SHOCK
Meets MIL-E-4970A (20 g's, 11 mSec in three principal axes).
TEMPERATURE COEFFICIENT OF OUTPUT
Less than 20 ppm per 'C from +10'C to +40'C